Download User Manual
Transcript
Calibration 51 Calibration of the pressure measurement We recommend an examination of the calibration once a year. Operate the function key pressure in the adjust-menu. The calibration of the pressure measurement works like described for the temperature calibration at up to six points. After the input of two points you have the choice whether to terminate the calibration (function stop input), or to enter further values (function another input). The pressure probe is applied with known pressure, and then the values will be entered successively via the numeric keyboard and confirmed with >ENTER<. The calibration points for the measurement on dialysis machines should be at -400, -100, 0, +100 and +500 mmHg. If you call the function stop input after your input, the entered values will be indicated once more, and can be taken over with the ok-key.