APPENDIX B
Elekta Neuromag MEG sensors
B.1 Sensor types
The detection coil geometry in MaxFilter is similar to the one described in
NM20568A Source Modelling Software User’s Guide Appendix B: “Coil
geometry information”.
The flux is an integral of the magnetic field component normal to the coil
plane. Thus, the basis matrix elements S nm for b in are approximated by:
in
S nm
Nk
=
∑ wkp νnm ( r kp ) ⋅ nkp
p=1
where r kp are a set of N k integration points covering the pickup coil
loops of the sensor, ν nm ( r kp ) is the value of the inside VSH function at
r kp , n kp are the coil normal directions at these points, and w kp are the
weights associated to the integration points. This formula essentially corresponds to numerical integration of the magnetic field over the pickup
loops of sensor k . The VSH terms for b out are approximated in a similar
way.
Table B.1 lists the parameters of the coil geometry descriptions employed
in the software. The colums of the table contain the following data:
1. The number identifying the coil type. This number is used in the coil
descriptions found in the FIFF files.
2. Description of the coil.
3. Number of integration points.
4. The locations of the integration points in coil coordinates.
5. Weights assigned to the field values at the integration points.
NM21993A
2006-10-31