Download MaxFilter 2.0 Manual
Transcript
B Elekta Neuromag MEG sensors Type Nk Description rk wk 2 Neuromag-122 planar gradiometer 8 ( ± 5.44, ± 7.68, 0 )mm ( ± 11.1, ± 7.68, 0 )mm ± 1 ⁄ ( 4 ⋅ 16.2mm ) 3012 Type T1 planar gradiometer Type T2 planar gradiometer Type T3 planar gradiometer 8 ( ± 5.89, ± 6.71, 0.3 )mm ( ± 10.8, ± 6.71, 0.3 )mm ± 1 ⁄ ( 4 ⋅ 16.69mm ) Type T1 magnetometer Type T2 magnetometer 9 ( 0.0, 0.0, 0.3 )mm Type T3 magnetometer 9 3013 3014 3022 3023 3024 ( ± 9.99, ± 9.99, 0.3 )mm ( 0.00, ± 9.99, 0.3 )mm ( ± 9.99, 0.0, 0.3 )mm 16 ⁄ 81 25 ⁄ 324 10 ⁄ 81 10 ⁄ 81 ( 0.0, 0.0, 0.3 )mm ( ± 8.13, ± 8.13, 0.3 )mm ( 0.0, ± 8.13, 0.3 )mm ( ± 8.13, 0.0, 0.3 )mm 16 ⁄ 81 25 ⁄ 324 10 ⁄ 81 10 ⁄ 81 Table B.1 MEG coil descriptions. B.2 Scaling between magnetometers and gradiometers The Elekta Neuromag® data contains both magnetometer and planar gradiometer channels which have different SI units (T and T/m, respectively). Therefore, some scaling between the channel data and basis functions has to be applied before combining them for Maxwell filtering. MaxFilter utilizes the inherent RMS noise levels of the sensors: n m = 3 fT/sqrt(Hz) for magnetometers, n g = 3 fT/cm/sqrt(Hz) for gradiometers. The basic modelling equation ( b = Sx , see Appenix A.2), is presented in the form diag ( w )b = diag ( w )Sx where the weight for the kth channel is w k = n g ⁄ n k . Thus, the gradiometer channel weight becomes 1, while the magnetometer channel weight is 100. Optionally, the weight w k for magnetometers can be set with the command-line option -magscale mult; the gradiometer weight is still one. 50 2006-10-31 NM21993A