4.3
Measurement Procedures and Tutorial
4.3.3 Ending Measurement
Auto Restart
The Auto Restart function is used to shut down the instrument and then restart it at a
specified time. Selections are made for the date and time that measurement will next be
performed, as well as for the instrument restart settings. After shutting down the
instrument, the carrier gas flow and power to the furnace are automatically turned off.
Approximately one hour before the restart time, the furnace and the carrier gas will
automatically be turned on to allow the instrument to achieve a Ready state by the restart
time.
1.
2.
3.
Select Auto Restart in the Standby window.
Select the appropriate instrument restart settings (TC furnace, carrier gas, ozone
generator) and set the restart date and time. If a check is placed in the box beside a
unit, the furnace power remains on and the carrier gas continues to flow. In order to
have the power turned off automatically, do not place a check mark in any of these
boxes.
Click the Standby button.
Power to the instrument is automatically switched off and will be automatically
switched on approximately one hour before the restart date and time. Using Auto
Restart, measurement can be performed immediately at the specified startup time.
Reference: Refer to Section 4.4.6.3 "Standby" for detailed descriptions of the available instrument
shutdown options.