Impact of Changing Instrument Settings Parameters Relationship to Grid Voltage% Delay Time and Grid Voltage% are interactive parameters. For each Grid Voltage% there is an optimum Delay Time, and for each Delay Time there is an optimum Grid Voltage%. The best approach for optimizing Delay Time is to leave the Grid Voltage% at a fixed value, and optimize Delay Time until you obtain optimum resolution. For more information, see Section 5.4.3.4, Optimizing Delay Time. Adjusting In general: • Higher masses require a longer Delay Time • Delay Time is affected by matrix For more information on Delay Time, see: • Section 5.4.3.4, Optimizing Delay Time • “Delay Time (nsec)” on page 5-18 5 Voyager™ Biospectrometry™ Workstation User’s Guide 5-55