Download E1460A 64-Channel Relay Multiplexer Module User`s Manual
Transcript
Scanning Channels Scanning the multiplexer module channels consists of closing bank channel(s) to the respective bank COM terminal(s) one channel at a time. Single scan, multiple scans (2 to 32767), or continuous scanning modes are available. Scanning Channels Comments Scanning Channels Sequence. The TRIGger:SOURce command specifies the source to advance the scan. The OUTPut command can be used to enable the E1406A Command Module "Trig Out" port, TTL Trigger bus lines (0-7), or ECL Trigger bus lines (0-1). Figure 2-5 illustrates scanning. ARM:COUN <number> TRIG:SOUR BUS|HOLD|EXT|MM| TTLT|ECLT OUTP ECLT|EXT|TTLT OUTP 1|0|ON|OFF INIT:CONT 1|0|ON|OFF SCAN:MODE NONE|VOLT|RES|FRES SCAN:PORT NONE|ABUS SCAN <channel_list> INIT TRIG Figure 2-5. Scanning Channels Sequence Chapter 2 Using the Relay Multiplexer 43