Download 5 - Sorensen
Transcript
Analog Programming (APG) and Isolated Analog Programming (ISOL) Resistive-Controlled Current ISOL Setup Activating ISOL Resistive-Controlled Current Setup 1. Turn the 9-position mode control to the CAP position or press the rotary Adjust/Enter control if the control knob is already at the CAP position. is displayed on the output voltage display. 2. Turn rotary Adjust/Enter control to select source from isolated connector). (voltage 3. Press the rotary Adjust/Enter control to commit the selected mode. is displayed on the output voltage display. 4. Select the input source level using the rotary Adjust/Enter control. 5. Once the desired level has been selected, press the rotary Adjust/ Enter control to commit the change. The display will return to the normal display. Deactivating ISOL Current Mode To deactivate the ISOL current mode: 1. 2. 4-30 Turn the 9-position mode control to the CAP position or press the rotary Adjust/Enter control if the control knob is already at the CAP position. is displayed on the output voltage display. 3. Turn the rotary Adjust/Enter control to select . 4. Press the rotary Adjust/Enter control to commit the change. The display will return to the normal display. M370430-01 Rev C