Download 4075B Series Arbitrary Function Generator User Manual
Transcript
This query is used to read the SCPI version to which the instrument complies. Command Type: Query Syntax; Response: Query only :SYSTem:VERSion? 1992.0 (NR2 format) Security :SYSTem:SECurity[:STATe] <Boolean> This command enables the instrument memory to be cleared. The stored settings and the arbitrary waveform memory are cleared when the Security state is changed from ON to OFF. The instrument state is returned to the factory power-on default. Arguments Type: Command Type: Setting Syntax: Examples: Query Syntax: Response: Boolean Setting or Query :SYSTem:SECurity[:STATe]<ws>ON|1|OFF|0 :SYST:SEC ON :SYST:SEC OFF :SYSTem:SECurity[:STATe]? 0|1 Power-on Buffer :SYSTem:POBuffer <buffer number> This command is used to set the Power On Buffer setting. The instrument will power-on with the setting stored in that buffer. Arguments Type: Range: Rounding: Command Type: Setting Syntax: Example: Query Syntax: Response: Numeric 0 to 49 to integer value Setting or Query :SYSTem:POBuffer<ws><buffer>|MINimum|MAXimum :SYST:POB 49 :SYSTem:POBuffer?[<ws>MINimum|MAXimum] Power-on buffer in NR1 format 86