MI 3002 & MI3102
Measurements
5.9 TRMS current (MI 3102 only)
This function enables measuring of AC currents in a wide range from 0.5 mA to 20 A
using sensitive current clamp (A 1018) supplied by METREL.
For additional information concerning TRMS current measurement refer to the
METREL’s handbook Measurements on electric installations in theory and practice.
How to perform a TRMS current measurement
Step 1 Select the TRMS CURRENT function (TRMS) with the function selector
switch. The following menu is displayed:
Figure 5.42: TRMS current measurement menu
Step 2 Set the following limit value:
‰
High limit current value.
Step 3 Connect the test leads to the Eurotest instrument and follow the connection
diagram shown in Figure 5.43 to perform a TRMS current measurement. Use
the Help function if necessary.
L1
L2
L3
N
PE
Option A 1018
IL
I
c
Figure 5.43: Connection diagram