Download ACQUITY UPLC System Operator`s Guide
Transcript
5. Slide the shelf into the sample organizer until it stops. Caution: To ensure the transfer shuttle moves freely and without damaging the sample organizer, you must be able to slide a shelf/plate/vial combination in or out without interfering with the restrictors directly above and below it. Shelf/plate combination and restrictors Sample plate Restrictors Shelf 6. Repeat steps 2 through 5 for the remaining plates. 7. Close the sample organizer door. A mechanism on the door ensures the shelves are positioned correctly when the door closes. Caution: To avoid jarring the plates from their shelves, do not slam the sample organizer door closed. 8. Click Configure > Scan and store shelf layout to update and save the new shelf configuration. To remove and replace the same plate on the same shelf: 1. Open the sample organizer door, and then remove plates that have finished processing. 2. Pull a shelf toward you, and then insert a plate of the same type and size on the shelf. 3. Load a plate onto a shelf so that position A,1 is at the right-rear corner and the forward edge of the plate is behind the stop at the left-front Preparing the sample organizer 3-31
Related documents
MonoTrap(溶媒抽出用)カタログPDF
Ultra Start 1250 SERIES Technical data
ACQUITY UPLC H-Class System Guide
User Manual for Electronic Emissions Reporting Software in
Waters 2996 PDA Detector
(PDA) detectors
Vials Chapter, 2010-2011 Waters Quality Parts®, Chromatography
2 - Waters
User Manual for Electronic Emissions Reporting Software in
Chap0 - Peak Parking Kit
XPEP Exosome Mass Spec Kit
U.P. Electronics Corporation Limited