Download ACQUITY UPLC System Operator`s Guide
Transcript
PDA detector alarm messages (Continued) Message Resolution HW Communications Failure, PDA Reference ADC 1. Check the Ethernet cable connection at the PDA detector and network switch. Reconnect the cable. 2. If the alarm persists, reset the PDA detector via the control panel or ACQUITY UPLC Console. 3. If the alarm persists, power-off and power-on again. 4. If the alarm persists, call Waters Technical Service (see page 7-4). HW Communications 1. Check the Ethernet cable connection at the Failure, PDA Sample ADC PDA detector and network switch. Reconnect the cable. 2. If the alarm persists, reset the PDA detector via the control panel or ACQUITY UPLC Console. 3. If the alarm persists, power-off and power-on again. 4. If the alarm persists, call Waters Technical Service (see page 7-4). HW Communications Failure, PDA Temp ADC 7-120 Diagnostics and Troubleshooting 1. Check the Ethernet cable connection at the PDA detector and network switch. Reconnect the cable. 2. If the alarm persists, reset the PDA detector via the control panel or ACQUITY UPLC Console. 3. If the alarm persists, power-off and power-on again. 4. If the alarm persists, call Waters Technical Service (see page 7-4).
Related documents
MonoTrap(溶媒抽出用)カタログPDF
Ultra Start 1250 SERIES Technical data
ACQUITY UPLC H-Class System Guide
User Manual for Electronic Emissions Reporting Software in
Waters 2996 PDA Detector
(PDA) detectors
Vials Chapter, 2010-2011 Waters Quality Parts®, Chromatography
2 - Waters
User Manual for Electronic Emissions Reporting Software in
Chap0 - Peak Parking Kit
XPEP Exosome Mass Spec Kit
U.P. Electronics Corporation Limited