Download Manual
Transcript
ToughSonic ® CHEM Family Sensors – Installation & Operating Instructions ToughSonic® LVL Family Sensors - Installation & Operating Instructions Mechanical Details Installation Dimensions are inches [mm]. Sensor measurements are the distance between the ultrasonic transducer face (the bottom of housing shown below) and the target material. The lower threads vary by model: Precautions 1.5-in Parallel: LVL-100 and LVL-300 2.0-in Parallel: LVL-500 Keep unintended targets from the transducer’s field of view. Keep the beam pattern in mind. Keep the transducer away from ultrasonic noise sources, such as pressurized air nozzles. Do not allow material to build up on the sensor face or sensor performance may suffer. Orientation Max. Width Orient the sensor perpendicular to the liquid surface or target object for best results as shown in Figure 3. Upper Threads 1-in. NPT Length Lower Threads 2-in. parallel or 1.5-in. parallel Figure 3 - Sensor Orientation Face 1.5 in. parallel: 1.9 [47mm] 2.0 in. parallel: 2.3 [58mm] Figure 2 – Housing Dimensions Ultrasound energy must reflect back to the sensor or the sensor will not detect the target, and may detect a later multipath echo (which would “measure” as a lower tank level). Location and Obstacles The sensor mounting location should be chosen so there are no obstacles in the beam path that reflect the ultrasound beam back to the sensor. Senix Corporation, 10516 Route 116 Suite 300, Hinesburg, VT 05461 USA 802.489.7300 or 800.677.3649, FAX: 802.489.7400, Web: www.senix.com, e-mail: [email protected] Page 9 of 38 – May 31, 2015