Download DynaCool Manual

Transcript
Section C.4
Cryostat Temperature Control (JQP-7)
C.4
Appendix C
Pin-out and Interconnect Diagrams
Cryostat Temperature Control (JQP-7)
This connector is located under the cryostat cosmetic covers, on the top plate of the cryostat to the left of
the sample chamber when facing DynaCool from the front. The “Serial Com” connections are used to
read thermometer calibration data, magnet configuration data, and other cryostat configuration
information. See Figure C-2 for connector diagrams.
Table C-3. JQP-7Cryostat Temperature Control.
C-4
PIN
FUNCTION
1
First Stage Thermometer I+
16
First Stage Thermometer I-
2
First Stage Thermometer V+
17
First Stage Thermometer V-
3
Second Stage Thermometer I+
18
Second Stage Thermometer I-
4
Second Stage Thermometer V+
19
Second Stage Thermometer V-
5
Magnet Thermometer I+
20
Magnet Thermometer I-
6
Magnet Thermometer V+
21
Magnet Thermometer V-
31
First Stage Heater I+
32
First Stage Heater I-
33
Second Stage Heater I+
34
Second Stage Heater I-
37
Cryopump Regeneration Heater I-
38
Cryopump Regeneration Heater I+
39
Impedance Heater I+
40
Impedance Heater I-
41
Pot Heater I+
42
Pot Heater I-
14
Serial Com Data
15
Serial Com +5 V
29
Serial Com Ground
30
Serial Com Clock
PPMS DynaCool User’s Manual, 1307-110, Rev. A0
November 2011
Quantum Design