Download DynaCool Manual

Transcript
Chapter 4
Theory of Operation
Section 4.3
Chamber Temperature Control System
Temperature. Some options, such as the Vibrating Sample Magnetometer, include a control
thermometer. In this case the Control Temperature is the value from this thermometer (e.g. the coil
thermometer for Vibrating Sample Magnetometer). This Control Temperature is reported in MultiVu
Log Data as “Control Temp (K)”. Also, the Log Data item “Control Therm (ID)” reports an
identification number for the control thermometer. A value of 643 indicates the block thermometer.
Other values indicate that Option Software has configured the system to use a different control
thermometer.
Sample Temperature. This value reports the temperature that is being controlled to achieve the desired
sample temperature. If no extended-range temperature option is active, this is the same as the Control
Temperature. If an extended range temperature option such as Dilution Refrigerator is active, then this
value is measured by an option-provided sample thermometer that is being controlled using a feedback
loop implemented in Option Software, not base DynaCool temperature software. This value is reported
in the MultiVu status bar in Log Data as “Temp (K)”. Also, the Log Data item “Sample Therm (ID)”
reports an identification number for the control thermometer. A value of 643 indicates the block
thermometer. Other values indicate that Option Software has configured the system to use a different
sample thermometer.
Measurement Temperature. This value is the temperature of the sample recorded in option data files
with sample measurement results. In many cases, this value is the same as the Sample Temperature, and
may be derived by averaging multiple readings of the Sample Temperature. In other cases, the
Measurement Temperature is derived from option thermometer readings not displayed elsewhere. An
example is Heat Capacity, where the Measurement Temperature is the average of the platform
thermometer during the relaxation measurement.
4.3.4.2 TEMPERATURE CONTROL CONFIGURATIONS
DynaCool provides a configurable temperature control platform to allow a measurement option to
control on a thermometer close to the sample and/or extend the temperature control range. The following
control configurations are available
Base Configuration. This configuration is used when no measurement option is active, or when the
measurement option provides no additional thermometers. In this mode, the Measurement Temperature,
Sample Temperature, Control Temperature, and Block Temperature all report the value of the block
thermometer. Measurement configurations that use this mode include:
Measurement
Configuration
Electro-Transport
or Resistivity
(1.8 – 400 K)
User
measurements
with no additional
thermometer
Control
Thermometer
Sample
Thermometer
Measurement
Thermometer
Control
Therm
(ID)
Sample
Therm
(ID)
Block
Thermometer
Block
Thermometer
Block
Thermometer
643
643
Block
Thermometer
Block
Thermometer
Block
Thermometer
643
643
Base Configuration with Measurement Option Thermometer. This configuration is used when the
measurement option provides a thermometer close to the sample that is used to measure the sample
temperature, but the control system does not use this thermometer. In this mode, the Sample
Temperature, Control Temperature, and Block Temperature all report the value of the block
thermometer. The Measurement Temperature reports the value of the measurement option thermometer.
Measurement configurations that use this mode include:
Quantum Design
PPMS DynaCool User’s Manual, 1307-110, Rev. A0
November 2011
4-9