Averaging averages values on an interval, which can be specified with the horizontal bar.
Smoothing performs a mobile average on the specified number of points, as entered in the field Points.
2. Press OK to confirm.
 Deleting one or more steps
This function may be useful if one or more steps were not correctly stored (i.e. data was recorded before starting the test,
signal gaps occurred, patient removed the mask).
Place the cursor on the step to be deleted and press Delete step to delete all data from the row.
 Advanced editing
By pressing Advanced you may perform advanced editing on the numerical values.
Mode
Function
Delete steps
Deletes the steps which meet the selection criteria.
Smoothing
Applies a moving average on the number of steps indicated in the Steps (smoothing) field, or on a
parameter as specified in the Edit parameter section.
Edit parameter
Edits a parameter according to the criteria in the Edit parameter section.
Edit parameter
Specifications
Parameter
Allows the parameter to be edited.
Value
Replaces the value of the selected parameter with the new value for all steps occurring in the time
range.
Correction %
Applies a percentage correction to the value of the selected parameter.
Offset
Adds an offset to the value of the selected parameter.
Formula
Replaces the value of the selected parameter with the results of a mathematical function entered
(see Formula editor).
Time range
Specifications
From, To
Specifies the time range for editing.