3. Valve outlet
Note: be very careful in order to differentiate inlet from outlet. These two are not
interchangeable, to guarantee proper functionality.
 Connecting the two-way non rebreathing valve to the mask
Depending on the turbine used (new one or old one), the following schemes apply:
A
1
2
3
4
B
9
6
5
7
8
Config. A (new turbine)
Config. B (old turbine)
1. VO2 max mask with adapter for turbine
5. VO2 max mask with adapter for old turbine
2. ID28 Turbine
6. Turbine
3. Adapter for 2-way valve
7. Optoelectronic reader
4. 2-way valve
8. Adapter for 2-way valve
9. 2-way valve
Please be careful to firmly connect all the parts.
Close the sampling line connector on the reader (turbine or optoelectronic reader) by means of the little plug attached to
the 2-way valve.
 Connecting the above assembly to the head support
 Patient’s preparation
Fix the mask as illustrated in the picture below. Adjust the elastic bands on the head cap as necessary to eliminate
possible leaks and create a tight seal around the subject’s face.
The mixing chamber - The mixing chamber- 259