Model 2700 Multimeter/Switch System User’s Manual Getting Started 1-15 Power-up Line power connection Follow the procedure below to connect the Model 2700 to line power and turn on the instrument. 1. Check to see that the line voltage indicated in the window of the fuse holder assembly (Figure 1-3) is correct for the operating voltage in your area. If not, refer to “Setting line voltage and replacing fuse,” page 1-16. CAUTION 2. Operating the instrument on an incorrect line voltage may cause damage to the instrument, possibly voiding the warranty. Before plugging in the power cord, make sure that the front panel power switch is in the off (O) position. Figure 1-3 Power module Model 2700 WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY. DIGITAL I/O TRIG. LINK ! RS232 MADE IN U.S.A. IEEE-488 ! SLT 1 KEITHLEY SLOT COVER SLT 2 Fuse Line Voltage Selector CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING. 100 220 240 120 Spring Window Fuse Holder Assembly