Download User`s Manual

Transcript
Model 2700 Multimeter/Switch System User’s Manual
Close/Open Switching Module Channels
2-19
OPEN key
The OPEN key has two options to open channels: ALL and MULTI. The ALL option
simply opens all channels in the mainframe. The MULTI option opens only the specified
channel. All other closed channels remain closed. Figure 2-7 summarizes OPEN key
operation.
OPEN: ALL — Perform the following steps to open all channels in the mainframe:
1.
2.
Press the OPEN key to display “OPEN: ALL.”
Press OPEN again (or press ENTER) to open all channels.
OPEN: MULTI — Perform the following steps to open only the specified channel:
1.
2.
3.
4.
5.
NOTE
Press the OPEN key. The “OPEN: ALL” message will be displayed.
Press the or key to display the “OPEN: MULTI” message.
Press ENTER to display the prompt to open a channel (OPEN MLT:XXX).
Using , , , and , key in the three-digit channel you want to select.
Press ENTER to open the channel.
If the channel you open using OPEN: MULTI is the system channel (channel
number displayed on the Model 2700), the channel will open, but the system
channel number will still be displayed (see “Multiple channel operation
anomalies,” page 2-22).
Figure 2-7
Multiple channel operation — opening one or all channels
OPEN CLOSE Press OPEN key
Display ALL option
and press OPEN again
OPEN: ALL
OPEN: MULTI
OPEN MULT:XXX
Display MULTI option
and press ENTER
Specify channel number
(XXX) and press ENTER