FIELD / WELL
MEASUREMENTS
Zero the probe as described earlier. For best results zero the probe at or
as close to the temperature of the sample as possible. Use the black
zeroing tube. Full five minute zeroing is recommended for best field work.
•
•
•
•
•
•
Lower the probe in the well or sampling area (close opening)
Observe the vapor measurements as you go down to your
desired depth.
If the readings are extremely high (≥10,000 ppm) it is likely that
there is product floating on the water. Do not allow probe to
contact pure product. If this occurs immediately, remove probe,
clean, precondition and calibrate it.
Continue using the Continuous mode until the probe enters the
water table. (This is indicated by an audible beep.) Exit the
continuous mode and use the 5 minute single measurement
mode to measure product concentration in water.
After measurement raise the probe back up and clean the probe
with one full wash bottle of zero water.
Store the (cleaned) probe in the blue humidity tube.
992041, Release 1.3
February 4, 2005
25