Download AmpFlSTR Sinofiler PCR Amplification Kit User Guide (PN 4384256C)
Transcript
Chapter 5 Experiments and Results Artifacts Artifacts and anomalies are seen in all molecular biological systems. Artifacts are typically reproducible, and anomalies are nonreproducible, intermittent occurrences that are not consistently observed in a system, for example, spikes and baseline noise. Artifacts have been seen in data produced on genetic analyzers when using the Sinofiler kit. In amplified samples, artifacts in the noncalling region may appear in the blue (95–100 nt) dye. Low-level artifacts in the calling region may appear in the blue (118 nt), green (97, 120, and 189 nt), and black (95–100, 164 nt) dyes, depending on the sensitivity of the instrument. Figure 5-7 on page 5-27 shows examples of baseline noise and artifacts in an electropherogram while using the Sinofiler kit. Genotyping may result in the detection of these artifacts as off-ladder alleles, or “OL Alleles.” You should consider possible noise and artifacts when interpreting data. 5-26 AmpFlSTR® Sinofiler™ PCR Amplification Kit User Guide