Download AmpFlSTR Sinofiler PCR Amplification Kit User Guide (PN 4384256C)

Transcript
Accuracy, Precision, and Reproducibility
Table 5-1 Precision results of five runs (16 capillaries/run) of the
AmpFlSTR® Sinofiler™ Allelic Ladder (continued)
Applied Biosystems 3130xl Genetic Analyzer
Allele
Mean
Standard Deviation
D12S391
8
215.54–215.79
0.05–0.066
9
219.5–219.77
0.054–0.071
10
223.5–223.79
0.047–0.067
11
227.42–227.71
0.045–0.062
12
231.38–231.68
0.046–0.066
13
235.33–235.63
0.042–0.067
14
238.35–238.67
0.053–0.075
15
239.3–239.63
0.054–0.07
D13S317
8
216.67–217.02
0.06–0.07
9
220.64–221.01
0.07–0.087
10
224.64–225.01
0.07–0.095
11
228.63–229.01
0.058–0.087
12
232.7–233.11
0.05–0.08
13
236.6–237
0.074–0.087
14
240.54–240.95
0.067–0.088
15
244.53–244.97
0.063–0.093
D16S539
5
AmpFlSTR® Sinofiler™ PCR Amplification Kit User Guide
252.29–252.66
0.068–0.085
5-7