Precision Decade
5.3.2

model 1427
Calibrating
Set the first calibration point (resistance element).
Use the keys ▲ and ▼ to set the element.

Measure the resistance of the selected element.
Use a ohm-meter with appropriate accuracy in 4 and 2-wire wire connection.

Switch the keypad to numerical mode.
To do this, press the “ENT” key.

Insert the measured resistance value in Ohms (Ω).
You’ll find the original value in the upper row and the new value in the lower row.

Confirm the new calibration data by pressing the “ENT” key.

Repeat the procedure described above for all resistance elements.