Download Precision Decade model 1427 OPERATION MANUAL
Transcript
Precision Decade 5.3.2 model 1427 Calibrating Set the first calibration point (resistance element). Use the keys ▲ and ▼ to set the element. Measure the resistance of the selected element. Use a ohm-meter with appropriate accuracy in 4 and 2-wire wire connection. Switch the keypad to numerical mode. To do this, press the “ENT” key. Insert the measured resistance value in Ohms (Ω). You’ll find the original value in the upper row and the new value in the lower row. Confirm the new calibration data by pressing the “ENT” key. Repeat the procedure described above for all resistance elements. Page 26