<READY>Message in alternately and Flash <HOLD> symbol ,if the weight inside the zero 'dead' band. 4. Place the mass to be weighed on the scale. 5. Once the weight moves outside the zero 'dead' band and stable signal, the instrument begins to calculate a long term average (selectable time, AVGTM) that compensates for any movement in the mass. During the selectable time the signal must be stable, if not, time starts again.The instrument flashes the <HOLD> symbol and shows the current average value. 6. The <HOLD> symbol is steady when the final sample weight is shown on the display. 7. Press the < HOLD > key to force the sample to be re-calculated. Press any other key to back to weighing Mode. 8. Once the weight is returned to the zero 'dead' band, the cycle can be repeated without press any key. * zero 'dead' band define is 5 Min.weights.(5 DD or 5 scale divisions) 54