Single Channel ECG Test System | User Manual
Power supply
Environment
Safety, EMC standards
USB +5Vdc supply (no
Typical load<0.25A, up
separate power supply
to 0.45A is possible if all
required)
relays are turned on
0.5A (high power mode)
15 ~ 30°C (by design, not Selection of components
tested)
is such that no effect
30 ~ 80% RH (design not from the environment is
tested)
expected.
No applicable safety standards (maximum internal
voltages 12Vdc)
For EMC no testing performed. CE marking based on
careful selection of parts, including USB protection
IC, as well as special filters to reduce noise from
microprocessor (8MHz) and DC/DC converter
(200kHz). Due to the low volume of production,
testing is not warranted.
Additional specifications may be provided on request.
1.5
Revision history
2011-11-23
2011-09-26
2011-08-11
2012-09-07
2012-07-19
2012-08-29
2014-01-01
Software upgrade for the following:
 Correct setting of T-wave amplitude during Auto Pacing
mode (previous software set this to 0.0mV, should be 0.2mV)
 Correct mode selection (due to addition of ANSI/AAMI
exponential waveform in the main function list, some
automated features selected the incorrect mode).
Software upgrade for the following:

Added functions for ANSI/AAMI EC 13
Software upgrade for the following:
 Reduction in streaming interruptions
 Software controlled calibration procedure
 Added new software functions for IEC 60601-2-27:2011
(heart rate testing)
 Updated software functions to better suit IEC 60601-2-51
Editorial corrections; add new section for options
Updated for SECG 4.0
Corrected editorial errors (change pictures, diagrams to SECG
4.0), and added section on terminals P1, P2 and P6
Re-layout and produce in production line