PASCAL® Firmware Release 4.0
Introduction
As of December 1, 2006, new PASCAL Tonometers are being shipped with upgraded firmware.
Release 4.0 incorporates some useful enhanced capabilities, while an effort has been made to keep
standard tonometer operations essentially unchanged relative to previous release 3.x firmware. This
preliminary document describes the new features. For all standard operations, please refer to the
PASCAL User Manual, version 2.1, which remains valid in all essential aspects unless specifically
mentioned in the present document.
The attached Firmware Block Diagram substitutes the previous diagram in Figure 16 (page 28) of
the vers. 2.1 User Manual.
Changes
Tonometer Start-up:
PASCAL now starts more promptly after activating it with a click on the Blue Knob. Particularly, restarting data acquisition while the unit is still active is now instantaneous.
Sound volume:
New PASCAL systems, starting with serial number 061140300 (S2 systems) or 061130330 (S1
systems) incorporate a beefed-up loudspeaker. Setting the speaker volume to 7 will now produce a
sound that will be easily audible even in a noisy environment. If too loud, set the volume to a lower
figure as described in the Manual (section 5.2.1).
User Settings:
The User Settings Menu (activated with a Long Click) has been reorganized and contains several
new features. For an overview, see the attached Block Diagram. Several menu items requiring
activated wireless hardware, will only be functional in wireless systems. Less frequently used
functions (e.g. volume, contrast, and language settings) have been moved to the end of the menu.
Special Modes:
All system test and special measuring modes (performance test, cantilever test, and manometer
mode) have been grouped under the "Special Modes" menu item. A new mode "Low OPA" has
been added and will be discussed in the "new features" section below.
PASCAL v40 firmware features.doc / 06-11-30 / 1
Ziemer Group AG
Allmendstrasse 11 / CH-2562 Port, Switzerland / PHONE: +41 32 332 7050
E-MAIL: [email protected] / www.ziemergroup.com
/
FAX: +41 32 332 7071
New Features
Q Indicator
PASCAL now monitors the quality of an incoming signal in real time while the user is taking data. As
soon as sufficient data have been accumulated to ensure a result with adequate quality can be
computed, the system will emit an audible signal. At that time the user can pull the tonometer off the
patient's eye, performing an Interrupt Maneuver as discussed in section 1.3 of the User Manual. In
the User Settings menu, click forward to the "Q Indicator" menu item, then go to the desired Quality
level (1,2,3,4), or set to "inactive" and allow the PASCAL to power off.
Example: If the Q Indicator is set to 3, and if you collect data until the special alert signal sounds,
you will obtain an IOP/OPA reading with Quality rating of 3 or better. (For a discussion of the Quality
rating, see User Manual section 8.2).
The factory setting of the Q Indicator is "inactive"; i.e. the PASCAL will function as pre-4.0 units.
The user decides how long to stay on the patient's eye to collect data. Remember: the longer the
data stream, the higher the chance that the program will find good quality data and produce a
highly reliable result. To get better results, increase the measurement time. We recommend a
measurement time of five seconds for optimum results.
Transfer Limit
Some users with PASCAL systems connected to a printer or a PC, find it unnecessary and timeconsuming to have all results transferring automatically to their external device. Setting the Transfer
Limit will suppress printing or PC transfer of any results that are below a pre-set range of Quality
scores.
Example: If the Transfer Limit is set to "Q = 1 - 3", then only measurements with a Quality of 3 or
better will print. Data with a poor quality rating of 4 or 5 will not print. These measurements will
however be retained in the Result List (numerical result only) for later viewing. Caution: the raw data
(Pressure curve) of any measurements suppressed with the Transfer Limit setting will be
irretrievably lost.
The factory setting for the Transfer Limit is "Transfer all" (i.e. no data transfers will be suppressed).
Low OPA Special Mode
In its normal operating mode, PASCAL requires the presence of a detectable OPA (approximately
0.5 mmHg or more). If no OPA is detected, an IOP cannot be computed. In special situations, you
may need to measure IOP on eyes that exhibit no measurable OPA (e.g. eyes with gas or silicon oil
tamponade, hypotonic eyes etc). In such situations, set your PASCAL to "Low OPA" Mode. This
setting can be found in the "Special Modes" menu.
Caution: In "Low OPA" mode, the firmware is not checking for periodicity in the data and could
therefore erroneously interpret bad data as acceptable, leading to an erroneous IOP reading. It is
recommended to visually inspect pressure curves recorded in Low OPA mode, using the optional
PASCAL printer or PASCAL DataWizard software, and to convince yourself that the data you have
recorded are plausible.
Attachment on following page: PASCAL Firmware Release 4.0 Block Diagram
PASCAL v40 firmware features.doc / 06-11-30 / 2
Ziemer Group AG
Allmendstrasse 11 / CH-2562 Port, Switzerland / PHONE: +41 32 332 7050
E-MAIL: [email protected] / www.ziemergroup.com
/
FAX: +41 32 332 7071
PASCAL® Firmware Release 4.0 Block Diagram
Menu Item
Values
(off)
L
s
s
"Please Wait. System Test"
"OK (recording)"
= Long Click
= Short Click
L
s
(last Î previous Î previous Î …)
Result List
L
s
s
inactive
Print History
activated
L
s
Special Modes
s
inactive
s
Low OPA
Performance Test
s
L
Cantilever
s
Data Transfer
Manometer
s
inactive
print values
s
print val/range
s
L
to PC
s
Wireless Setup
print all
s
inactive
Printer search
s
PC search
s
L
Firmw. download
s
Transfer Limit
s
Transfer all
Q=1-4
s
Q=1-3
s
L
Q=1
s
Q Indicator
Q=1-2
s
Q=4
Q=3
s
Q=2
s
L
inactive
s
System Info
Firmw.Rel.#
s
Q=1
Tip S/N
s
L
Key Value
s
Sys. S/N
Batt. L/N
s
s
Batt. P/N
s
Sys. P/N
Tip P/N
s
00 Î 01 Î 02 Î 03 Î 04 Î 05 Î 06 Î 07
Volume
L
s
01 Î 02 Î 03 Î 04 Î 05 Î 06 Î 07 Î 08
Contrast
L
s
Language
English Î Deutsch Î Français Î Italiano Î Español Î Portugues
© 2006 SMT AG
061129 awi