Download Operation Manual
Transcript
Enter menu P1 Calibration. Scroll to parameter 04 Temp. calibr. Press the navigation key. 8.1.4. Calibration test During calibration test value of calibration mass is compared to present result of weighing. It is just checking and the result have no influence on changes of factors. After displaying results press key ESC. Calibration test a) b) c) d) e) Move to submenu P1 – Calibration. Scroll to function Calibration test Press key Right The instrument makes test Check results Cal. – load on the pan Act. – calibration mass in moisture analyzer memory Dif. – difference between these two values Return to weighing mode in point 6 of instruction manual. 8.1.5. Temperature calibration Parameter enebling to carry out temperature sensor calibration in drying chamber. 8.1.6. Printout of the calibration report After each calibration report can be printed. The report is printed on external printer or sent to computer and recorded in the file to archivize. P1 08 Print report : 1 : on – report is printed P1 08 Print report : 0 : off – without printing report - 21 -