C-Link Protocol Commands Measurement Configuration Table B-6. Operating Ranges, continued Selection ppm mg/m3 8 5,000 7,500 9 C1 C1 10 C2 C2 11 C3 C3 custom range range = | 1 | 2 | 3 | This command reports the user-defined value of custom range 1, 2, or 3. The example below reports that custom range 1 is defined to 55.0 ppm. Send: Receive: custom 1 custom 1 5500E-2 ppm set custom range value set custom 1 value set custom 2 value set custom 3 value set custom 1 range value set custom 2 range value set custom 3 range value These commands are used to set the maximum concentration for any of the three custom ranges 1, 2, or 3 to range value, where value is a floating-point number representing concentration in ppm or mg/m3. The example below sets the custom 1 range to 55.5 ppm. Send: Receive: set custom 1 range 55.5 set custom 1 range 55.5 ok range mode This command reports the current range mode. Send: Receive: range mode range mode single set range mode mode This command sets the current range mode to single, dual, or auto. The example below sets the range mode to single. Send: Receive: B-30 Model 42i High Level Instruction Manual set range mode single set range mode single ok Thermo Fisher Scientific