Download User manual - Mars

Transcript
МС2.211.001 UM
ƒ amplitude ratio of the harmonic (“Un” or “In” field, the value is set as a percentage of
the 1st harmonic):
ƒ “Interharmonics” mode is off. Values are set within the range of 0 to 100.00 % in
0.01 % increments.
ƒ “Interharmonics” mode is on. Values are set within the range of 0 to 15.00 % in
0.01 % increments.
ƒ relative phase of the harmonic (“ϕ1–n”, the value is set as a percentage of the 1st harmonic within the range of –179.99° to 180.00° in 0.01° increments);
ƒ symbolic name of the channel (to switch over 6 channels).
The “Spectrum (linear)” option (Fig. 4.11) represents spectrogram of the signal in the
channel under review drawn in linear scale.
Fig. 4.11. “Waveshape” window
with linear spectrum of amplitudes
The “Spectrum (logarithmic)” option (Fig. 4.12) displays the spectrogram of the signal in
the channel under review drawn in logarithmic scale (in decibels, dB), this only affects graphical part of the window. The value in its modifiable field named “amplitude ratio of the harmonic” (Un or In) are still displayed and edited in linear scale (as a percentage of the 1st harmonic).
Fig. 4.12. “Waveshape” window
with logarithmic spectrum of amplitudes
The “Harmonics Phase Shift” option (Fig. 4.13) displays the spectrum of phase shift angles, each determined relative to the first (fundamental) harmonic of the signal in the active
channel.
19