Viscosity vs. Shear Rate: Displays all of the results of the viscosity measured at all the shear rates.
Data and Report Saving: At the end of execution, user will be prompted
for data and report saving. In the prompt, the location of temporary report file in html will be disclosed and the same report will be printed to a
default printer. Click OK to save a copy of the files and print.
Note: Using the VROC_SP program, completely fill the VROC chip flow
channel with test sample (no bubbles inside the channel) before making the
baseline measurement for the evaluation test. This is a very important step
to ensure accurate baseline measurements. A partially filled channel will
give an erroneous reading of baseline. If there is liquid inside the chip,
make sure that at least 80 µl of sample is pumped to the chip to replace
the liquid with the test sample. If there is enough sample, 150 µl elution
volume is needed.