Download IQ-MSD Turbo Software User Manual
Transcript
IQ–MSD Turbo 1.4 Advanced IQ System Software Osc Frequency: Sets the frequency in hertz for the above function. It does not affect the generator when a test is in progress. Sweep The sweep parameters define the operation of the generator during a normal test sweep. Note: If all SLM-8s have the same generator settings, a single generator can be used for all of them. Resolution: Sets the frequency step size of the test sweeps. It can be set to 1⁄24, 1⁄12, 1⁄6, 1⁄3, 1⁄2, 2⁄3 and 1 octave steps. Less resolution (higher numbers) will result in faster sweeps with less detail if the Sync parameter is turned off. Higher resolution (smaller fractions) will result in slower sweeps with more detail. If multiple units are synchronized, their Resolution settings must match. Start Freq: Sets the frequency at which the generator will start the test sweeps. Although you can set this parameter in 1 Hz steps, the actual starting frequency will be equal to the closest octave setting as determined by the previous Resolution parameter. The lowest possible setting is 20 Hz and the Start Freq setting should always be lower than the Stop Freq setting. If multiple units are synchronized, their Start Freq settings must match. Stop Freq: Sets the frequency at which the generator will stop the test sweeps. Although you can set this parameter in 1 Hz steps, the actual stopping frequency will be equal to the closest octave setting as determined by the previous Resolution parameter. The highest possible setting is 22000 Hz (22 kHz) and the Stop Freq setting should always be higher than the Start Freq setting. If multiple units are synchronized, their Stop Freq settings must match. Sync: Turn on this control if multiple units are being synchronized together. Turn it off if this is not the case. When turned on, it forces the test start frequency, duration and stop frequency to be the equivalent of an eightchannel test regardless of the number of channels actually being used. When turned off, the duration of a test is determined by the number of channels being used in the one unit. Note: The previous sweep parameters must be set the same for all units when Sync is turned on. Test The following parameters configure other aspects of a test: Impedance Tolerance: Sets the criteria by which impedance test data is evaluated to determine if the load should pass or fail a test. It simply compares the impedance test data to the reference impedance data to see how much they differ. If the difference is greater than the Impedance Tolerance setting, the load has failed the test. The range is ±1 to ±100%. Frequency Tolerance: Sets the criteria by which frequency test data is evaluated to determine if the output should pass or fail a test. It simply compares the frequency magnitude test data to the reference frequency magnitude data to see how much they differ. If the difference is greater than the Frequency Tolerance setting, the output has failed the test. The range is ±1 to ±25 dB. Page 199