5
Programming Command Reference
At start up of the Tester, the queries in <int 1> and <int 2>
format are returned.
If the setting with <int> format is performed, the format of the
query changes to <int> format.
If the setting with <int 1> and <int 2> format is performed,
the format of the query returns to <int 1> and <int 2>
format.
This query command returns the set Band and Channel number
of RFCH with integer. The range of the value is same as that of
the command parameter.
Frequency
TXANalyzer:RFCH:FREQuency <real>
TXANalyzer:RFCH:FREQuency?
This command sets the frequency. The allowable range and
preset values are as follows:
Table 5-19 Frequency Allowable Range and Preset Value
5-110
Band
Allowable Range [Hz]
Preset Value
1
1920000000 to 1980000000 in 200000 step
1922400000
2
1850000000 to 1910000000 in 200000 step,
1852500000, 1857500000, 1862500000,
1867500000, 1872500000, 1877500000,
1882500000, 1887500000, 1892500000,
1897500000, 1902500000, 1907500000
1852400000
3
1710000000 to 1785000000 in 200000 step
1712400000
4
1710000000 to 1755000000 in 200000 step,
1712500000, 1717500000, 1722500000,
1727500000, 1732500000, 1737500000,
1742500000, 1747500000, 1752500000
1712400000
5
824000000 to 849000000 in 200000 step,
826500000, 827500000, 831500000,
832500000, 837500000, 842500000
826400000
6
830000000 to 840000000 in 200000 step,
832500000, 837500000
4162
N9360A Multi UE Tester W-CDMA Programming Manual