3.4.3
Calibration Preparation
The following is a general outline of the steps required to calibrate all the sensors:
" Select a calibration standard whose value is near that of your field samples.
" Remove the Storage Cup from the Transmitter.
" Clean and prepare the sensors as detailed in Sections 3.4.4 through 3.4.8.
" Attach the Calibration Cup.
" Using the Calibration Cap, thoroughly rinse the sensors several times by half-filling the
calibration cup with deionized water and shaking the Transmitter to make sure each sensor
is free from contaminants that might alter your calibration standard.
"
In a similar manner, rinse the sensors twice with a small portion of the calibration
standard, each time discarding the rinse.
19