Operation Instrument Controls Menu VALVE POWER: >ACTUATOR BLOWBACK DILUTION AIR PITOT BLOWBACK TEOM BYPASS SAMPLE PUMP RELAY BYPASS PUMP RELAY CFACT INST ON OFF ON ON OFF ON ONª DIAGS ALARM PI PD PS Set Temperatures OFF OFF OFF The Set Temperatures menu allows the user to view and set all thermistor user desired values. ● In the Main Menu, choose Instrument Controls > Set Temperatures. SET TEMPERATURES: >DILUTION STINGER BYPASS NEPH BLOCK TEOM AIR TUBE TEOM CASE FAILSAFE CFACT INST Set Temperatures o 165 120 100 58 62 65 0 C C C o C o C o C o C o o DIAGS ALARM The Set Temperatures screen allows the user to view and set temperature parameter for the selected component. The example below shows the Dilution Temp screen. The stinger, bypass, neph block, TEOM air tube, TEOM case, and failsafe screens function the same way. ● In the Main Menu, choose Instrument Controls > Set Temperatures > (temperature). SET DILUTION TEMP: CURRENTLY: 165 SET TO: 160 o o C C ? ©ª INC/DEC SAVE VALUE CFACT INST Set Flows 3-14 Model 3880i Instruction Manual DIAGS ALARM The Set Flows menu allows the user to view and set user desired flow values. Thermo Fisher Scientific