Download Ion Max APPI Source User Manual Version C
Transcript
5 Optimizing the LTQ MS Detector with Your Analyte Setting Up the MS Detector for APCI/APPI/MS Operation Figure 30. Define Scan dialog box, showing typical settings for APCI/APPI/MS operation g. In the Source Fragmentation group box, confirm that the On check box is not selected ( ) to specify that the ion source fragmentation option is turned off. h. In the Scan Ranges group box, in the Input list box, select Center/Width to make available the Center Mass and Width text boxes in the Scan Ranges table. i. In the Scan Ranges table, in the Center Mass text box, enter 609.20 to set the center mass for the scan range to m/z 609.20. j. In the Width text box, enter 2.00 to set the width of the scan range to m/z 2.00. k. Ensure that the settings in your Define Scan dialog box are the same as those shown in Figure 30. l. Click OK to apply the MS detector scan parameters and to close the Define Scan dialog box. 5. On the Control/Scan Mode toolbar, click Centroid/Profile to toggle the data type to centroid. The picture on the button should be the same as that shown here. 6. Click Positive/Negative to toggle the ion polarity mode to positive. The picture on the button should be the same as that shown here. You have now completed setting up your MS detector for APCI/APPI/MS operation. Thermo Scientific APPI Souce User Manual 57