Remote Operation 4-31 Device Action — Figure 4-12 provides a detailed look at the device action. If the repeat filter is enabled, then the instrument samples the specified number of reading conversions to yield a single filtered reading. If the moving filter is active, or filter is disabled, then only one reading conversion is performed. Figure 4-12 Device action (trigger model) From Delay block of Trigger Model (See Figure 4-11) To Output Trigger block of Trigger Model (See Figure 4-11). Device Action Conv Conv Conv Hold Chan Filtering Process (Filter enabled) Conv = Reading conversion Hold = Hold Feature process (if enabled) Chan = Close channel (if scanning) If the hold feature is enabled (see :HOLD commands in Section 5), then the first processed reading becomes the "seed" reading and operation loops back to the beginning of the device action. After the next reading is processed, it is compared to the programmed hold window (0.01% to 20%). If the reading is within the window, then operation again loops back to the beginning of the device action. This looping action continues until the specified number (2 to 100) of valid hold readings (readings within the window) have occurred. If one of the hold readings is not within the window, then the instrument acquires a new "seed" reading and repeats the hold process. After the hold is released, an audible beep is sounded to signal a valid measurement. The use of Hold is explained in Section 3. If the instrument is performing a step or scan, then the next task for device action is to open the previous channel (if closed) and close the next channel. If the filter, hold feature and scanning are disabled, the device action would simply be a single reading conversion.