Download Model 2000 Multimeter
Transcript
2-28 Basic Measurements MX + B This math operation lets you manipulate normal display readings (X) mathematically according to the following calculation: Y= mX + b where: X is the normal display reading m and b are user-entered constants for scale factor and offset Y is the displayed result Configuration To configure the mX+b calculation, perform the following steps: 1. Press SHIFT then MX+B to display the present scale factor: M: +1.000000 ^ 2. 3. Enter a value and units prefix. Use the and keys to choose a numerical place and use the ▲ and ▼ keys to increment or decrement the digits. Press ENTER to confirm the M value and display the B value: B: +00.00000 m 4. 5. Enter a value and units prefix. Press ENTER to confirm the B value and display the UNITS designation: MXB 6. Scroll through the letters to change and press ENTER when done. The Model 2000 then displays the result of the calculation. Percent This item selects the percentage calculation and lets you specify a reference value. The displayed reading will be expressed as a percent deviation from the reference value. The percentage calculation is performed as follows: Input - Reference Percent = ------------------------------------------ × 100% Reference where: Input is the normal display reading. Reference is the user entered constant. Percent is the displayed result. Configuration To configure the percent calculation, perform the following steps: 1. Press SHIFT then % to display the present value: REF:+1.000000^