Download Model 2000 Multimeter
Transcript
Specifications A-8 Optimizing measurement accuracy The configurations listed below assume that the multimeter has had factory setups restored. DC voltage, DC current, and resistance: • • • Select 6½ digits, 10 PLC, filter ON (up to 100 readings), fixed range. Use REL on DC voltage and 2-wire resistance measurements. Use 4-wire resistance measurements for best accuracy. AC voltage and AC current: • Select 6½ digits, 10 PLC, filter ON (up to 100 readings), fixed range. Temperature: • Select 6½ digits, 10 PLC, filter ON (up to 100 readings). Optimizing measurement speed The configurations listed below assume that the multimeter has had factory setups restored. DC voltage, DC current, and resistance: Select 3½ digits, 0.01 PLC, filter OFF, fixed range. AC voltage and AC current: Select 3½ digits, 0.01 PLC, filter OFF, fixed range. Temperature: • Select 3½ digits, 0.01 PLC, filter OFF. For all functions, turn off the display and autozero and set the trigger delay to zero. Use the :SAMPle:COUNt and READ? bus commands.