To create an injection list: (continued)
Step
Action
4
The injection list is automatically filled in with information from the selected Sample
Sheet.
Fill in the module information from the Module drop-down list.
5
See “Selecting the Module and Dye Set/Primer File” on page 6-19 if you need help
selecting the correct file.
Note
Drag the column markers at the top of columns to change their width.
Fields in the sequencing injection list form:
Field
Description
Sample Sheet
Displays the selected sample sheet name.
Page Icon
Opens the selected sample sheet.
Length to Detector
The distance between the end of the capillary that is in the
sample and the capillary window. The length you enter will not
affect the configuration of the instrument. It is for record-keeping
purposes only.
Operator
The name you enter here will appear on printed data.
Tube & Sample Name
The sample’s name and position in the autosampler.
Module
The file that contains the specific functions executed to process
samples.
Inj. Secs
The duration of the injection in seconds.
Inj. kV
The voltage during the injection in kilovolts.
Run kV
The voltage during the run in kilovolts.
Run °C
The temperature of the heat plate during the run in degrees
Celsius.
Run Time
The duration of the data collection in minutes. This is less than
the total run time.
Finish Time
Displays the time at which raw data collection for the sample
finished.
Note
You may need to scroll to see the Run Time and Finish Time columns.