Failures/troubleshooting
10 Failures/troubleshooting
10.1 Volume flow functional circuit
10.1.1 Blower does not run up after
switching the sampler on.
Possible causes:
1. Failure power electronics:
When reaching the range end, the failure indication status
"Blower overloaded" is displayed.
Cause:
Failure in the power electronics (frequency convertor) and/or the
blower. Control electronics is operational.
Failure elimination:
Remove the compartment cover of the suction unit/frequency
convertor.
a) Check whether protection logics of the frequency convertor
has responded. The corresponding failure indication
message appears on display of the frequency
convertor.With blowers of a negligible running time (1 to 2
years), after resetting the frequency convertor, there can be
immediately tried a new sampler start-up using the main
switch of the instrument. Mind keeping the lower
compartment still open. Upon running up, check the blower
for suspect noise generation, while the display of the
frequency convertor has to show the output voltage
frequency (210 Hz as a maximum).
b) With older blowers (more than 2 years), first check the blower
carefully before re-starting.
c) If the frequency convertor gets into the failure indication
status immediately after trying to switch the sampler on and
this occurs also with the hooked off blower, the frequency
convertor is defective. This unit should be sent to the factory
to get repaired (resp. replaced).
2. Possible operating error:
Time program is not in the "Work” status.
blower is switched off manually
Sampler is in "Remote control" mode.
filter changer is not in sampling position
3. Heavy contamination:
The flow meter glass measurement tube is heavily
contaminated, e.g. after the sampler ran in a heavily exposed
environment without an inserted round filter for a longer time.
Please remove and clean the glass measurement tube.
4. Control system
a) The printed circuit board (PCB) "flow control" is defective
(replace the board).
b)The infrared photo-sensors configuration on the flow meter
is defective (call the service).