Download Operation instructions DH77 Touchscreen
Transcript
Version HW0.12 Failures/troubleshooting 10 Failures/troubleshooting 10.1 Volume flow functional circuit 10.1.1 Blower does not run up after switching the sampler on. Possible causes: 1. Failure power electronics: When reaching the range end, the failure indication status "Blower overloaded" is displayed. Cause: Failure in the power electronics (frequency convertor) and/or the blower. Control electronics is operational. Failure elimination: Remove the compartment cover of the suction unit/frequency convertor. a) Check whether protection logics of the frequency convertor has responded. The corresponding failure indication message appears on display of the frequency convertor.With blowers of a negligible running time (1 to 2 years), after resetting the frequency convertor, there can be immediately tried a new sampler start-up using the main switch of the instrument. Mind keeping the lower compartment still open. Upon running up, check the blower for suspect noise generation, while the display of the frequency convertor has to show the output voltage frequency (210 Hz as a maximum). b) With older blowers (more than 2 years), first check the blower carefully before re-starting. c) If the frequency convertor gets into the failure indication status immediately after trying to switch the sampler on and this occurs also with the hooked off blower, the frequency convertor is defective. This unit should be sent to the factory to get repaired (resp. replaced). 2. Possible operating error: Time program is not in the "Work” status. blower is switched off manually Sampler is in "Remote control" mode. filter changer is not in sampling position 3. Heavy contamination: The flow meter glass measurement tube is heavily contaminated, e.g. after the sampler ran in a heavily exposed environment without an inserted round filter for a longer time. Please remove and clean the glass measurement tube. 4. Control system a) The printed circuit board (PCB) "flow control" is defective (replace the board). b)The infrared photo-sensors configuration on the flow meter is defective (call the service). 37 DIGITEL Elektronik AG Alte Gasse 18 CH-8604 Hegnau Tel.: +41 (44) 908 20 30 DIGITEL Elektronik GmbH Illstrasse 30 A-6706 Bürs Tel.: +43 (5552) 67850 www.digitel-ag.com [email protected]