Version HW0.12
Operation
(3)
QCal integrated flowmeter: actual set volume flow (actual conditions)
at the integrated flowmeter for volume flow of 500
l/min at the inlet under station conditions.
pintegrated flowmeter: actual pressure in the integrated flowmeter
during calibration
Tintegrated flowmeter: actual temperature in the integrated
flowmeter during calibration
QScale = Qloc × (Tref × ploc) (pref × Tloc)
For operation volume flow of 500 l/min under station conditions
the following conditions will be on the integrated flowmeter:
TStation flowmeter = TStation + 3K (approximate value)
pStation flowmeter = pStation – pfall at filter (will be meassured automatically
during calibration)
pStation integrated flowmeter: average air pressure at the installation site
minus pressure fall at filter at volume flow of 500 l/min. Which
means: the air pressure of measuring system if the air pressure
at the inlet is the same pressure as at the station.
TStation integrated flowmeter: average temperature at the installation site
plus 3K temperature increase at filter at volume flow of 500
l/min. Which means: the temperature of the measuring system if
the temperature at the inlet is the same temperature as at the
station.
from the general gas equation
Q1 ×
p1
p2
= Q2 ×
T1
T2
it
or
QCal build - in flowmeter = QRef ×
(7)
Tbuild - in flowmeter × TStation build - in flowmeter
pbuild - in flowmeter × pStation build - in flowmeter
From the general gas equation and from the equation (7) follows
the volume flow on calibrated flowmeter:
follows:
QCal Ref = QRef ×
Qloc buil - in flowmeter (Station) =
pStation
TStation + 3K
×
=
(4) QRef ×
TStation pStation - pfall at filter
pStation × TStation build - in flowmeter
QRef ×
TStation × pStation build - in flowmeter
QRef:
Qloc indoor Station:
pStation:
TStation:
QScala build - in flowmeter = QRef ×
(5)
pStation
×
TStation
pbuild - in flowmeter × TStation build - in flowmeter
Tbuild - in flowmeter × pStation build - in flowmeter
QCal Ref: flow on calibrated flowmeter (under actual
conditions), so that reaching the operation volume
flow of 500 l/min (at the inlet) under station
conditions.
Pcal. flowmeter: actual air pressure in the calibrated flowmeter
(during calibration is the same pressure at the
separator)
Tcal. flowmeter: actual air temperature in the calibrated flowmeter
(during calibration is the same temperature at the
separator)
From the equation (3) and the equation (8) follows the flow to be
set on calibrated flowmeter (under actual conditions), thus
reaching the operation volume flow of 500 l/min (at the inlet)
under station conditions.
(9)
pStation
×
TStation
pbuild - in flowmeter× TStationbuild - in flowmeter× Tcal.flowmeter× Tref
Tbuild - in flowmeter× pStationbuild - in flowmeter× pcal.flowmeter× pref
QScaleCalRef = QRef ×
Tref × TStation build - in flowmeter
pref × pStation build - in flowmeter
QScala integrated flowmeter: Shown volume flow at integrated
flowmeter under station conditions at inlet for air inlet
volume flow of 500 l/min. This value is automatically
taken over by the control software as a set flow.
At these settings the volume flow at the inlet is QRef (500 l/min) if
the station conditions are given. The volume flow for different
conditions during calibration follows from equation (2) and (5):
QCal build - in flowmeter = QScala build - in flowmeter ×
(6)
pStation × Tcal. flowmeter
×
TStation × pcal. flowmeter
(8)
Air inlet volume flow of 500 l/min. under
station conditions.
volume flow in the integrated flowmeter for
air inlet volume flow of 500 l/min. under
station conditions.
average air pressure at the installation site
average temperature at the installation site
At the station conditions at the integrated flowmeter and from
equation (3) and (4) follows:
pStation
×
TStation
pref × Tbuild - in flowmeter
Tref × pbuild - in flowmeter
QScale cal Ref: flow to be set on calibrated flowmeter, so that
reaching the operation volume flow of 500 l/min (at
the inlet) under station conditions.
7.4.3
Error estimates
A frequent question emerges how errors in temperature or
pressure measurements or deviation from assumption applied to
the determination of the operation volume affect the calculated
standard resp. operation volumes. The order of magnitude of
these errors is illustrated below using several examples. Further,
there are also stated affects of deviations of actual station
conditions during the sampling period how the entered station
conditions affect upon calibration of the instrument.