Download Horizon Programming Manual
Transcript
Basic Programming and Testing Enter the minimum resistance measurement for the Isolation test in kΩ. Select Enter to continue. To restore the Quick Scan setting, simply select Isolation Min Res again. Select Conductor Res to select a measured value for the Continuity measurement. Select the desired Continuity Type. For this exercise, select Resistance @5mA. Resistance @5mA MSR measured scan. Quick Scan As described above. HC Resistance Programmable HCS measured scan. (Resistance reference) Voltage Drop Programmable HCS measured scan. (Voltage reference) From Sample Uses the MSR to measure the maximum resistance for the attached nets. Set the Maximum Conductor Resistance accordingly. Fixed HC Measured scan using the Fixed HCS. (Voltage or resistance reference). The Fixed HC option can only be used for 4-wire measurements. Enter the maximum conductor resistance (measurements made higher than the value entered will be considered opens). Select Enter to continue. Horizon Series Tester: Programmer’s Manual 4-17