Download 3172-W2 VXIbus Arbitrary Waveform Generator User Manual

Transcript
3172 User Manual
Publication Number 980949 Rev. H
Table 7-2, Calibration for W2 Module
7-10
Item in
Cal Panel
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
DMM
Function
VDC
VDC
VDC
VDC
VDC
VDC
VDC
VDC
VDC
VDC
VDC
VDC
VDC
VDC
VDC
VDC
VDC
VDC
VDC
VDC
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
VAC
VAC
VAC
VAC
VAC
VAC
VAC
VAC
VAC
VAC
VAC
VAC
VAC
VAC
VAC
VAC
VAC
VAC
Low Limit
Target
High Limit
(Volts)
(Volts)
(Volts)
-0.020
0.00000
0.020
-0.020
0.00000
0.020
-0.020
0.00000
0.020
-0.020
0.00000
0.020
-0.020
0.00000
0.020
-0.050
0.00000
0.050
0.995
1
1.005
2.490
2.5
2.510
3.980
4
4.020
5.475
5.5
5.525
6.965
7
7.035
8.460
8.5
8.540
9.460
9.5
9.540
-1.005
-1
-0.995
-2.510
-2.5
-2.490
-4.020
-4
-3.980
-5.525
-5.5
-5.475
-7.035
-7
-6.965
-8.540
-8.5
-8.460
-9.540
-9.5
-9.460
Set DMM function to Volts AC
1.7528
1.7678
1.7828
0.52533
0.53033
0.53533
0.17528
0.17678
0.17828
0.052533
0.053033
0.053533
0.017178
0.017678
0.018178
0.0048033 0.0053033
0.0058033
1.7528
1.7678
1.7828
0.52533
0.53033
0.53533
0.17528
0.17678
0.17828
0.052533
0.053033
0.053533
0.017178
0.017678
0.018178
0.0048033 0.0053033
0.0058033
1.7528
1.7678
1.7828
0.52533
0.53033
0.53533
0.17528
0.17678
0.17828
0.052533
0.053033
0.053533
0.017178
0.017678
0.018178
0.0048033 0.0053033
0.0058033
END
Adjust Pot
RV10
RV2
Astronics Test Systems