Download Triggering the Digitizers
Transcript
Table 2-1. Power-on and Reset States. Parameter Power-on/Reset State Parameter Power-on/Reset State OUTPut:TTLT0-7:STATe OFF (all TTLTrigger lines) TRIGger:LEVel2 -256 V (Channel 2 level) ROSCillator:SOURce INTernal TRIGger:LEVel3 -256 V (Channel 3 level) SWEep:POINts 1 (one sample) TRIGger:LEVel4 -256 V (Channel 4 level) SWEep:OFFSet:POINts 0 (no pretrigger samples) TRIGger:SOURce1 IMMediate (source 1 not Ch 1) VOLT1:RANGe 256 V (Channel 1 range) TRIGger:SOURce2 HOLD (source 2 not Ch 2) VOLT2:RANGe 256 V (Channel 2 range) TRIGger:SLOPe1 POSitive (slope 1 not Ch 1) VOLT3:RANGe 256 V (Channel 3 range) TRIGger:SLOPe2 POSitive (slope 2 not Ch 2) Input Overload Condition Overload voltages may occur which will open the channel input relay disconnecting the input signal from the channel. Overload voltage by range is shown in the following table. Range Voltage Input Condition Vmax 62 mV to 4 V High or Low to Guard >20 V 16 V to 256 V Low to Guard >40 V The overload is reported both when the readings are retrieved and when the next measurement is initiated. If an overload occurred, an error message is returned when data is retrieved informing you that the data is questionable (Overload detected - data questionable). An error message is also returned when you initiate the next measurement (Overload detected - attempting re-connect of input relays). NOTE Relays open at approxiately 260 V. If this happens, you must reprogram the input range to close by executing INP <channel> ON. 36 Using the Digitizers Chapter 2