Download Manual - Frye Electronics
Transcript
The HEARLab ACA Measurements 29 6. For CTE assessments, you must also specify the stimulus transducer used to present the stimuli. Do this by selecting the transducer from the Stim. Transducer list (all the calibrated transducers of the transducer type selected from Stim. Presentation will be shown) 7. For ACA assessments, you may need to specify the loudspeaker position (if there is more than one loudspeaker set up). Do this by selecting the position from the Loudspeaker Position list. 8. Select the masking type from the Masking to use list: • None • Narrow Band Noise (only available for CTE) 9. Where applicable, specify the masking presentation from the Masking Presentation list: • Insert Earphone • Bone Conductor 10. Where applicable, specify the calibrated masking transducer that is to be used to present the masking from the Masking Transducer list. 11. Where applicable, select the masking presentation level relative to the stimulus level from the Masking Level list: -40 dB, -30 dB, -20 dB, -10 dB, 0 dB, +10 dB. 12. Acquisition of cortical responses will stop automatically when the stop criterion is reached. This stop criterion is based on the minimum number of accepted epochs for responses to each stimulus. Specify this number as shown below: 13. Click OK to proceed with assessment or Cancel if you do not wish to proceed. The New Run window (Figure ) will be displayed after OK is clicked.