Using the SAMPLEJET
Sample Holder Numbers
5.6
In TopSpin and IconNMR the samples are addressed by numbers. Newer versions of IconNMR will show the rack enumeration. For details see "SAMPLEJET
Together with IconNMR" on page 80
There are gaps within the list. They should not be used, otherwise the SAMPLEJET will return a „sample missing“ message.
The exact numbers of the holders is depending on the carousel. See the holder
numbers in the corresponding chapter of the carousel:
•
•
•
For carousel 99/5 read "Carousel 99/5" on page 65
For carousel 30/5 cooled read"Cooling (option)" on page 71
For carousel 47/5 read "Carousel 47/5" on page 77
Table 5.3.
The holder numbers
Previous IconNMR
sx commands
Newer IconNMR
1 - 30, 47 or 99
1 - 30, 47or 99
30, 47 or 99
Samples in the openshop ring.
101 - 196
1A1-1H12
96
For short samples in the rack 1
201 - 296
2A1-2H12
96
For short samples in the rack 2
301 - 396
3A1-3H12
96
For short samples in the rack 3
401 - 496
4A1-4H12
96
For short samples in the rack 4
501 - 596
5A1-5H12
96
For short samples in the rack 5
Number of slots
Used for
To set the sample position enumeration according to your needs, check the "User
Settings" on page 53. The default starts with A1, B1 ... H1 and jumps to A2.
40 (183)