5 Operation
To ensure that the current passes through the gel, all components of the transfer
sandwich are cut to the same size as the gels to be transferred.
40
8.
Saturate the cathode electrode plate with distilled water and remove any excess
with absorbent paper. Place the cathode on top of the transfer sandwich and
connect the socket on the black cathode lead to the cathode pin in the Multiphor II
base.
9.
Close the Multiphor II safety lid and connect the unit to the power supply. It is
recommended to run the transfer at a constant current of 0.8 mA/cm2. A transfer
time of approximately 1 hour is normal.
Note:
The current is calculated using the surface area (total length x width) of the
transfer sandwiches, and this calculation applies irrespective of the number of
transfer sandwiches in the stack.
Note:
For transfer times longer than one hour turn off the power supply, remove the
safety lid and carefully lift the cathode (top) electrode without disturbing the filter
papers or gel. Carefully pour on additional transfer buffer to re-wetting the filter
paper.
Multiphor ll Electrophoresis System User Manual 18-1103-43 Edition AK