Multi-Point Calibration
6.
Scroll to the right until you see the columns labeled Level 1, Level 2, etc.
7.
Enter the concentration of each peak in the standard mixture for each level in the
calibration curve.
8.
Level 1 should correspond to the lowest concentration.
9.
For the purpose of this tutorial enter 5 in the Level 1 column for each peak in the
Peak ID table.
10.
Enter 10 in the Level 2 column for each peak in the Peak ID table.
11.
Enter 20 in the Level 3 column for each peak in the Peak ID table.
Figure 88
12.
Peak ID with concentration information
entered.
To save the method with the new Peak Table information, click on the Save icon
on the Tool bar and select Save Method.
Calibration