Full-Pipe-Velocity Sensor Ground Button Velocity Electrodes The Model 2000 can be configured with either a one-inch full-pipe-velocity sensor or a two-inch full-pipe-velocity sensor (Figure 8). The installation instructions for the full-pipe sensors are contained in the manuals titled “One-Inch Full-Pipe Sensor Installation” and “Two-Inch Full-Pipe Sensor Installation. The sensor disconnect is required when the unit is configured with a full-pipe sensor. Figure 8. Full-Pipe Velocity Sensor SENSOR MOUNTING CONFIGURATIONS Universal-Sensor Mount The sensor can be attached to different size poles with the universal-sensor mount (Figure 9). Mounting instructions are as follows: • Insert the mounting shaft on the universal mount into the hole at the back of the sensor. The thumbscrew needs to be seated in the groove, so make sure the shaft is completely inserted into the hole. • Hand tighten the thumbscrew. • Slide a pole one inch or less in diameter through the clamp and tighten. Figure 9. Universal Sensor Mount [CAUTION] Do not over tighten the thumbscrew on the sensor. Excessive force on the thumbscrew could damage the sensor. 16