To Activate Chop Sweep Mode You can use the chop sweep mode to make two measurements at the same time. For example, the analyzer can measure A/R and B/R simultaneously. You can activate the chop mode by pressing 4PRESET5 or by the following the sequence below. NNNNNNNNNNNNNN NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN Press 4CAL5 MORE CHOP A and B . For more information, refer to \Alternate and Chop Sweep Modes" in Chapter 6. To Use External Calibration Ooading the error correction process to an external PC increases throughput on the network analyzer. This can be accomplished with remote only commands. Refer to the HP 8753D Programmer's Guide for information on how to use external calibration. To Use Fast 2-Port Calibration With the 2-port calibration on, faster measurements may be made by not measuring the reverse path for every forward sweep. This is controlled by the test set switch command. When making measurements using full two-port error-correction, the following types of test set switching can be dened by the user: Hold: In this mode the analyzer does not switch between the test ports on every sweep. The measurement stays on the active port after an initial cycling between the ports. The fastest measurements can be made by using this type of test set switching. Pressing the 4MEAS5 key or changing to a dierent S-parameter measurement will cause the test set to switch and cycle between the ports. Continuous: In this mode the analyzer will switch between the test ports on every sweep. Although this type of test set switching provides the greatest measurement accuracy, it also takes the longest amount of time. Number of Sweeps: In this mode there is an initial cycling between the test ports and then the measurement stays on the active port for a user-dened number of sweeps. After the specied number of sweeps have been executed, the analyzer switches between the test ports and begins the cycle again. This type of test set switching can provide improved measurement accuracy over the hold mode and faster measurement speeds than continuous mode. 1. To access the test set switch functions, press: 4CAL5 NNNNNNNNNNNNNN NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN MORE TEST SET SW 2. To activate the hold mode, press: 405 4x15 NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN The analyzer will then display TEST SET SW HOLD . 3. To activate the continuous mode, press: 415 4x15 NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN The analyzer will then display TEST SET SW CONTINUOUS . 5-54 Optimizing Measurement Results
Related documents
Agilent 8504B Precision Re ectometer
Agilent 4396B Network/Spectrum/Impedance Analyzer Function
Safety Summary - Silicon Integrated Circuit Reverse Engineering
HP 70004A - User Guide
8753ES Option 011 Network Analyzer User`s
Agilent 83480A User`s Manual
User Manual - Excalibur Engineering
About this Manual We`ve added this manual to the Agilent
Maxim Integrated MAX2057EVKIT datasheet
Cold Fusion User Manual
Agilent Technologies PN 8510-16 Programming instructions
HP 71910A - User`s Guide