Download GE Dash 2000 - User manual
Transcript
ECG: Electrode Placement 3-Leadwire Electrode Placement When a 5-leadwire electrode configuration is not desirable, a 3-leadwire electrode configuration can be used. 348A AHA Configuration 348B IEC Configuration NOTE: Electrode configuration will vary depending on the type of leadwire set you are using. Refer to ”Three-leadwire Configuration” below. Right arm and left arm electrodes should be placed just below the right and left clavicle. Left leg electrode should be placed on a non-muscular surface on the lower edge of the rib cage. Three-leadwire Configuration The molded 3-leadwire sets can be placed in the 5-lead Multi-Link patient cable. Standard Molded 3-leadwire Set (Rotating Reference) • Selectable lead I, II, or III cable with a rotating reference (right arm, left arm, left leg). Operation of the monitor with this standard cable allows you to select one of three leads (I, II, or III) for monitoring. When using the standard, 3-leadwire configuration, the following operating conditions occur: • • • Revision B LD ANALYSIS automatically switches to single lead analysis. If an attempt is made to change to multi-lead analysis, a message "NOT POSSIBLE WITH 3 LEAD CABLE" will appear briefly and no change will occur. DISPLAY LEAD choices are limited to I, II, and III. Any options usually allowing more than one ECG lead selection will be disallowed. A message line is displayed briefly, indicating such. Dash 2000 Patient Monitor 227 499 06 (US/E) 8-5